Low-to-High Temperature Chamber Model:Hot Disk TCUs
This floor-standing TCU uses a two-stage refrigeration system with R404A and R23 refrigerants for cooling, and a fan with a heating element around it for heating. It comes with a box-shaped temperature chamber with a volume of 50 liters.
-
The chamber has a chemically resistant stainless steel lining, accessed from the front side via a left-hinged lever latch door that is lockable. The sensor cable feedthrough is selectable either on the left side or the right side of the chamber via small circular wall openings having removeable rubber plug covers. The TCU is equipped with wheels, allowing for easy mobility in the laboratory. The temperature range of the chamber is impressive, down to -60°C, and up to 300°C. The temperature stability in turn is better than ±0.1°C at lower temperatures and better than ±0.5°C at higher temperatures.
The low-to-high temperature chamber is primarily intended to be used with our bare Kapton-insulated Hot Disk® or Hot Strip® sensors mounted in our sensor interface for high-temperature accessories, in combination with dedicated high-temperature sample holders.
The TCU is automatically controlled via the Hot Disk Desktop App over RS232 communication. The Hot Disk Desktop App can be programmed to automatically perform a series of measurements with the Hot Disk instrument at multiple chamber temperatures. -
Temperature Chamber
Size Volume 50 liters Height 440 mm Width 410 mm Depth 350 mm Cooling system Two-stage refrigeration system with R404A and R23 refrigerants for stage 1 and 2, respectively Heating system Electrical heating element with fan Temperature Minimum -60°C Maximum 300°C Stability ±0.1°C to ±0.5°C
GeneralSupply voltage 220-240 VAC, 50 Hz, 1 Phase Power rating 20 A (maximum) Input/Output RS232 connector (female) Dimensions
(height x width x depth)1440 mm x 785 mm x 865 mm Weight 177 kg CE marking CE markingIn progress